- Filter
MRI Filter F2780Y Serie
• CAT6a filter provides clean signal and data into Shielded rooms or MRI rooms through RJ45 Connector.
• Filters are compatible with IEEE Type 1/Type 2/Type 3/Type 4 PoE application up to 100W PSE
• The filters provide:
• 50dB attenuation from 30MHz to 70MHz.
• 100dB attenuation from 100MHz to 1GHz.
• 120dB attenuation from 1GHz to 10GHz.
• Voltage Rating: 100VDC
• Hi-Pot: Line to Ground: 200VDC for 2 sec
• Ambient Temperature Range: -25°C To +85°C
• Filter components are encapsulated in a plated metal enclosure
• The filters comply with UL and cUL construction requirements
• Filters are compatible with IEEE Type 1/Type 2/Type 3/Type 4 PoE application up to 100W PSE
• The filters provide:
• 50dB attenuation from 30MHz to 70MHz.
• 100dB attenuation from 100MHz to 1GHz.
• 120dB attenuation from 1GHz to 10GHz.
• Voltage Rating: 100VDC
• Hi-Pot: Line to Ground: 200VDC for 2 sec
• Ambient Temperature Range: -25°C To +85°C
• Filter components are encapsulated in a plated metal enclosure
• The filters comply with UL and cUL construction requirements
| Teilenummer | Description | Herunterladen |
|---|---|---|
| MRI Room signal line filter,1Amp, RJ45 connectors provided for input and output connections | Data Sheet |