• Filter

MRI Filter F3772Z 系列

• MRI Room three line filters deliver clean power to MRI scanning equipment to prevent interference with their performance. The filters are designed with minimum leakage current to ground and provide 100dB attenuation minimum from 5MHz to 10GHz.
• Voltage Rating: 120VAC/277VAC
• Frequency: 50/60Hz
• Hi-Pot: Line to Ground: 2250VDC for 2 sec, Line to Line: 1450VDC for 2 sec
• Ambient Temperature Range: -25°C To +85°C
• Filter components are encapsulated in a plated metal enclosure
• Leakage Current: 50mA max. @ 277VAC 60Hz
• UL and cUL listed
零件编号 Description 下载
MRI Room three lines filters, 50Amp, #6 AWG wire for output connections Data Sheet
MRI Room three lines filter, 30Amp, #8 AWG wire for output connections Data Sheet
MRI Room three lines filter, 15Amp, #10 AWG wire for output connections Data Sheet
Inquiry Cart

您的购物车中包含 0 products

样品购物车

您的购物车中包含 0 products

报价购物车

您的购物车中包含 0 products