• Filter

F2779Y001L 系列

• MRI Room data line filters provide clean communication lines into MRI rooms. The filters are designed with minimum leakage current to ground and provide 100dB attenuation from 1MHz to 10GHz, Pass Band 0-200KHz.
• Voltage Rating: 120
• 250V
• Frequency: DC to 60Hz
• Hi-Pot: Line to Ground: 1000VDC for 2 sec Line to Line: VDC for 2 sec
• Ambient Temperature Range: -25°C To +85°C
• Filter components are encapsulated in a plated metal enclosure
• Leakage Current: 5mA @ 120VAC 60Hz
• Comply with UL and cUL construction requirements
零件编号 Description 下载
MRI Room data line filter,1Amp, #22 AWG wire for output connections Data Sheet 3D Step File Data Sheet